Estimate shipping
Description
The LEEP PRECISION™ Integrated System brings together the three core pieces of a LEEP setup: the electrosurgical generator, the smoke evacuator and a mobile cart. It is built for gynecology clinics and offices that perform loop electrosurgical excision procedures and want one rolling workstation instead of separate components.
Features:
- Complete system: LEEP PRECISION generator, smoke evacuator and cart
- 120V model (order code LP-10-120)
- Mobile design moves easily between exam rooms
- Components are also sold individually (LP-20-120 generator, LP-30-120 smoke evacuator, LP-40-001 cart)
- Works with CooperSurgical LEEP electrodes and LEEP PRECISION disposables
Hardware warranty registration is available through CooperSurgical.
Package Includes:
- LEEP PRECISION Generator (120 V)
- LEEP PRECISION Smoke Evacuator (120 V)
- LEEP PRECISION Cart with integration unit
- Electrodes, pencils, return electrodes, filters and tubing sold separately
Specifications:
| Components | LEEP PRECISION generator, smoke evacuator, and cart with integration unit (single master ON/OFF powers both units) |
| System dimensions (W × D × H) | 22.44 × 22.07 × 41.22 in (57.0 × 56.06 × 104.7 cm) |
| System weight | Approx. 88 lb (40 kg) assembled |
| System electrical | 120 VAC, 60 Hz; fuses 10 A 250 V T-type slow-blow |
| RF output | 100 W rated, 495 kHz ± 5% into 500 ohms; monopolar only |
| Modes | CUT (continuous sinusoid, crest factor 1.2, 600 V max open circuit); BLEND (interrupted, crest factor 2.5, 600 V); COAG (short bursts, crest factor 6.5, 4000 V) |
| Power range | 10 to 100 W in every mode; accuracy ±20% above 10 W |
| Activation | Hand-switch pencil or foot pedal (not simultaneously) |
| Safety features | Patient return monitor cuts output if the dual (split) return electrode loosens; self-monitoring of all detectors; RF leakage 100 mA max |
| Duty cycle | Intermittent: 2 minutes on, 6 minutes off |
| Smoke evacuation | 35 CFM minimum; 3-stage filtration (pre-filter, charcoal odor filter, hydrophobic ULPA rated 99.999% at 0.014 micron) |
| Generator size / weight | 9.38 × 11.63 × 4.25 in (23.8 × 29.5 × 10.8 cm); 10 lb 9 oz (5.25 kg) |
| Smoke evacuator size / weight | 9 × 9 × 22 in (22.9 × 22.9 × 53.2 cm); 21 lb (9.5 kg) |
| Cart | Heavy-duty casters (two locking), two storage shelves, grip handle, cable hooks |
| Generator classification | Class I, Type BF defibrillator-proof applied part, IPX0 |
| Operating environment | 10 to 40 °C, 10 to 90% RH, 700 to 1060 hPa |
| Warranty | 1 year |